Review Article - Imaging in Medicine (2016) Volume 8, Issue 2
Catheters and guide wires for interventional MRI: are we there yet?
Helene CM Clogenson*1 & John J van den Dobbelsteen11Department of Biomechanical Engineering, Delft University of Technology, Mekelweg 2, 2628CD Delft, The Netherlands
- Corresponding Author:
- Helene CM Clogenson
Department of Biomechanical Engineering
Delft University of Technology, Mekelweg 2
2628CD Delft, The Netherlands
Tel: +31 (0) 15 278 95 15
E-mail: H.C.Clogenson@tudelft.nl
Abstract
Clinical feasibility of interventional MRI has been shown for a number of procedures. However, MRI-guided endovascular interventions have not yet reached clinical routine due to the lack of suitable MR safe or MR conditional and MR visible devices. In this article, some of the MR tools currently under development are reviewed. Good handling and MR visibility properties both are critical factors for successful application of such novel devices. Finally, we argue that it is only when both MR catheters and guidewires are available on the market that real progress and impact will be made in the field of interventional MRI.
Keywords
MRI; catheters; development; guidewires; interventional
Introduction
Catheterization in humans started in 1929, when, Forssmann [1] introduced a ureteral catheter into one of his left forearm veins. Following his pioneering work, new tools and techniques were rapidly developed. However, it was only by the end of the 40s and beginning of the 50s that critical steps forward were made, in particular with the development of both thin-walled [2] and radiopaque [3] catheters (soon to be braided for improved mechanical properties [4]) as well as suitable guidewires. The availability of these novel instruments led to safe minimally invasive techniques, as we know it today, with millions of patients annually diagnosed and treated worldwide.
Interventional radiology and cardiology have proven to be successful. Yet, they present several drawbacks. Above all, the interventions are performed under 2D projection fluoroscopy/ angiography guidance, which exposes both patient and medical staff to ionizing radiation. Today, we are witnessing yet another revolution in interventional radiology, based on the realtime magnetic resonance imaging. However, just like nearly 65 years ago, when new instruments were needed to take the first crucial steps forward in interventional radiology and cardiology, new MR suitable and visible catheters and guidewires are needed now to open the way for interventional MRI.
Magnetic resonance imaging (MRI), offers several advantages over conventional imaging modalities for both patients and interventionalists. MR imaging slices can be oriented in three dimensions [5,6], no ionizing radiation is used, and, unlike X-ray images, MRI has an excellent soft-tissue contrast [5-11] with no known harmful effects [5]. Additionally, MRI provides morphologic as well as functional information. With these additional features, MRI may help to determine the result of an endovascular intervention [8-10, 12] and provide safety advantages over conventional X-ray-guidance, for example by revealing more rapidly vascular perforation or rupture [12,13].
In the last 10 years, the feasibility of MRIguided vascular interventions has been demonstrated for a number of procedures in humans, such as angioplasty of femoral artery stenoses [14], iliac angioplasty and stenting [15], cardiac interventions [9,16,17], cardiopulmonary interventions [18], and others. Real-time MR sequences are now available [8] and most MR scanner manufacturers offer realtime MR sequences adequate to guide MRI catheterization [11]. Nevertheless, MRI-guided vascular interventions are not yet ready to be widely used in the clinical setting [19]. This is due to several issues linked to this imaging modality: increased operational complexity, reduced temporal and spatial resolution, restricted access to the patient and limited patient monitoring, and amount of noise during the scan [7-10,19,20].
For these reasons, critics argue that interventions performed under MRI-guidance represent a costly and cumbersome alternative to procedures that, otherwise, are conducted rapidly and efficiently under X-ray guidance and adjunctive ultrasound [12]. However, even if radiation exposure has been reduced over the years, X-ray guided procedures still pose significant health risks, particularly for children and young adults requiring multiple procedures, and staff with chronic exposure [9,11,12,21- 26]. Therefore, MRI-guided interventions might be justified. In order to reach clinical routine, the aforementioned mentioned issues need to be addressed and, most importantly, there is a need for suitable (MR safe or MR conditional) and visible MR instruments [5,7- 10,12,19,20,27,28]. Such devices being under development are reviewed, together with their mechanical handling and MR visualization properties.
MR Instruments
■Catheters and guide wires
Conventional devices designed for use under fluoroscopy-guidance cannot be used in MRI due to the frequent presence of ferromagnetic components (e.g., braiding in catheters) and/or of long conducting materials (e.g., guidewires) [9,27], which results in a risk of radiofrequency-induced heating [29,30]. Some commercially available catheters, stents, or balloons, though most not yet CE marked or FDA approved for MR-guidance, were used for MR-guided procedures, to a limited extent [9]. Indeed, MRI-guided interventions are still restricted by the lack of appropriate MRI instruments [28] with proper mechanical characteristics [5,31]. With conventional tools, proper pushability and torquability are achieved by the use of metallic cores (guidewires) and metallic braiding (catheters) [5], elements that have to be avoided, or limited in length, for the construction of MR instruments. As a result, MR instruments present usually lower bending and torsional stiffness and, therefore, reaching the target in the body may be difficult. As an example, non-braided angiographic catheters, used for cardiac catheterization under MRI-guidance, were described to have a low torque and poor steerability [9].
Consequently, nylon, aramid, and polyester (or other fibers), can be a solution for the construction of MR fiber-braided catheters with improved mechanical properties [32]. Examples of such tools can be found in the literature, e.g., with PEEK [33], nylon [34], or other [35] braids. The MR cardiac electrophysiology catheter developed on the basis of nylon braided tubing [34] was shown to have mechanical properties comparable to commercial MR incompatible catheters for cardiac catheterization. It seems that such fiber-reinforced catheters have the potential to become suitable catheters for MR interventions.
Nowadays, many MR catheter developments have been reported. Various groups or companies are getting interested in MR catheters for specific interventions such as cardiac electrophysiology (Imricor, Burnsville, USA; [16,33-35], biopsy catheters (ITP, Bochum, Germany), cardiac and renal ablation catheters (Imricor; [36]), cryoablation catheters [37], deflectable catheters (ITP, Bochum, Germany; [33,38]), as well as catheters for tracking [39-41], and many others. However, only a few of these catheters are suitable for use in the clinic and even less is available on the market. Also, to our knowledge, no MR visible catheters (and sheaths) suitable for MRI-guided endovascular navigation are CE marked or FDA, which makes the development of interventional MRI procedures more than challenging.
Till recently, the lack of MR guidewires has further limited the possibilities to perform MRIguided interventions, although some centers perform studies on patients without the use of a guidewire [11,16,42]. However, without guidewire support, successful navigation can be difficult or even impossible [11,16], or catheter shaft kinking may occur [11]; both incidents possibly leading to interventional failure.
In the development of MR devices, most of the efforts have initially focused on catheters [43]. In fact, it was only in 2012 that the first MR conditional and MRI visible endovascular guidewire received CE Mark (EPflex GmbH, Dettingen, Germany). Originally, conventional nitinol guidewires were used [44,45] but were found to be unsafe [46]. Then homemade PEEK [43,47] or other polymer-based [46] guidewires have been developed. However, as for catheters, the polymer-based guidewires had low bending stiffness [43], and fiber-reinforcedpolymer designs were introduced. For example, a 0.032” MR guidewire prototype was found to have improved mechanical properties compared to the polymer materials alone [20]. Another promising glass-fiber reinforced 0.035” guidewire [27] was developed and tested in several patients [17]. However, it was subsequently observed that with such glass-fiber reinforced tools, special precaution must be implemented in the design to prevent guidewire disruption during the intervention [48]. The possibility of guidewire disruption led to the development of a new composite version of the 0.035” MR guidewire [49] (MR conditional, MR visible, Fraunhofer Institute IPT, Aachen, Germany; Nano4imaging GmbH, Aachen, Germany). The 0.035” MR guidewire from EPflex is also such a composite design: PEEKbased with fibers [50]. Finally, a MR safe fiberreinforced MR guidewire portfolio (0.035” standard and stiff, 0.014”, and soon 0.018”), comprising the first continuous visualization of the guidewire shaft in MRI over its entire length, is currently awaiting CE Mark (MaRVis Medical GmbH, Hannover, Germany). Alongside these promising MR guidewires with passive-negative markers, another MR conditional guidewire with active marker is under development [51].
MR visualization
In parallel to the development of catheters and guidewires for MRI-guided interventions with proper mechanical properties, special care has to be taken in order to ensure that the developed MR instruments are visible on the MR images. Even when the instruments are suitable to be used in MR environment, they may present visualization issues, i.e. too big artifacts that shade the anatomy or, quite the reverse, they may not be visible on the MR images. It is known that polymer-only tools are not visible unless modified [5], or they may only be indirectly visualized.
Some catheters, for example, can be perceived on MR images using various methods: placement of an MR visible guidewire [45] in its lumen, or by injecting MR contrast agent [52], or just relying on the (potentially weak) susceptibility artifact created by the plastic [38]. In the same way, some balloon catheters can also be visualized by filling the balloon with air or with contrast media [11], or with carbon dioxide [9,42]. These tricks remind us the ways by which the early catheters were visualized on X- ray images before they were made radiopaque, e.g., with a metal guidewire inserted in the catheter or by filling the catheter with contrast medium [3,53,54].
Alternatively, markers or equipment can be added on the catheter or device in order to make it MRI visible. However, if a device is modified in such a way, the instrument loses its approval for human application. Even if such an approach is a solution for in-vitro and in-vivo experiments, it is a problem to get approval for a clinical study on patients. Therefore, special care needs to be taken during instrument development in order to ensure they offer appropriate MRI visualization properties. The devices can either be visualized using passive, semi-active, or active MR markers [5]. In addition, there is an increasing desire by the interventionists to visualize the entire instrument shaft and the exact tip localization for safe endovascular navigation [55,56].
Discussion
Extensive developments in interventional radiology (and cardiology) only happened when both suitable guidewires and catheters were available. Currently, the development of interventional MRI is restricted by the limited availability of these crucial tools. While the field has been hampered for a long time by the lack of MR guidewires, the recent developments are hopefully going to provide clinics with suitable and visible MR guidewires (EPflex, Nano4imaging, or MaRVis). Now, that such guidewires are being developed and approved, tests should be performed to evaluate if these MR guidewires have suitable mechanical properties for endovascular navigation and catheter support. To our knowledge, no such tests have been published yet.
Moreover, even now that some specialized catheters are being developed and are currently waiting for regulatory approval (e.g., Imricor), there is still only a limited number of MR catheters on the market. Some solutions have been found for clinical applications, but these are not always suitable for application in human. Most importantly, so far no navigation/exchange catheters, or introducers or sheaths specifically for use in interventional MR have been developed. These simple instruments must be developed to be both safe under MRI-guidance and visible on the images. Of course, to make MRI intervention a reality, these instruments should come at reasonable cost, especially for single-use navigation catheters and guidewires, which excludes complex designs.
It has been frequently reported that the main obstacle for adopting MRI catheterization is the unavailability of visible and MR safe/conditional endovascular devices [5]. On one hand, the recently developed MR conditional and MR safe guidewires allow for a larger array of MR-guided procedures to be investigated. However, on the other hand, it is only when both MR guidewires and catheters, visible, reliable, and with suitable mechanical properties, are developed that real progress and impact will be achieved in the field of interventional MR.
Disclaimer
H.C.M. Clogenson is currently working for the start-up company MaRVis Medical (GmbH, Hannover, Germany), which is developing MR-safe guide wires. The current manuscript was written as guest researcher at the Delft University of Technology, and it is based on H.C.M. Clogenson’s PhD work [57].
References
- Forssmann W. The catheterization of the right side of the heart. Klin. Wochenschr. 8, 2085– 2087 (1929).
- Higgs ZC, Macafee DA, Braithwaite BD et al. The Seldinger technique: 50 years on. Lancet. 366, 1407-1409 (2005).
- Odman P. Thoracic aortography by means of a radiopaque polythene catheter inserted percutaneously. Acta. radiol. 45, 117-124 (1956).
- Geddes LA, Geddes LE. The catheter introducers. Mobium Press, Chicago (1993).
- Saikus CE, Lederman RJ. Interventional cardiovascular magnetic resonance imaging: A new opportunity for image-guided interventions. JACC. Cardiovasc. Imaging. 2, 1321-1331 (2009).
- Krämer NA, Krüger S, Schmitz S et al. Preclinical evaluation of a novel fiber compound MR guidewire in vivo. Invest. Radiol. 44, 390-397 (2009).
- Saborowski O, Saeed M. An overview on the advances in cardiovascular interventional MR imaging. MAGMA. 20, 117-127 (2007).
- Kos S, Huegli R, Hofmann E et al. MRcompatible polyetheretherketone-based guide wire assisting MR-guided stenting of iliac and supraaortic arteries in swine: Feasibility study. Minim .Invasive. Ther. Allied.Technol. 18, 181–188 (2009).
- Razavi R, Hill DL, Keevil SF et al. Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. Lancet. 362, 1877-1882 (2003).
- Kos S, Huegli R, Hofmann E et al. First magnetic resonance imaging-guided aortic stenting and cava filter placement using a polyetheretherketone-based magnetic resonance imaging-compatible guidewire in swine: Proof of concept. Cardiovasc. Interv. Radiol. 32, 514–521 (2009).
- Ratnayaka K, Faranesh AZ, Hansen MS et al. Real-time MRI-guided right heart catheterization in adults using passive catheters. Eur Heart. J. 34, 380-389 (2013).
- Lederman RJ. Interventional Cardiovascular MRI. In: Kwong RY, editor. Contemporary Cardiology: Cardiovascular Magnetic Resonance Imaging. Humana press, Totowa 711–733 (2008).
- Raval AN, Telep JD, Guttman MA et al. Real-time magnetic resonance imaging-guided stenting of aortic coarctation with commercially available catheter devices in Swine. Circulation. 112, 699-706 (2005).
- Paetzel C, Zorger N, Bachthaler M et al. Feasibility of MR-guided angioplasty of femoral artery stenoses using real-time imaging and intraarterial contrast-enhanced MR angiography. Rofo. 176, 1232-1236 (2004).
- Manke C, Nitz WR, Djavidani B et al. MR imaging-guided stent placement in iliac arterial stenoses: A feasibility study. Radiology. 219, 527-534 (2001).
- Grothoff M, Piorkowski C, Eitel C et al. MR imaging-guided electrophysiological ablation studies in humans with passive catheter tracking: Initial results. Radiology. 271, 695- 702 (2014).
- Tzifa A, Krombach GA, Krämer N et al. Magnetic resonance-guided cardiac interventions using magnetic resonancecompatible devices: A preclinical study and first-in-man congenital interventions. Circ. Cardiovasc. Interv. 3, 585–592 (2010).
- Rogers T, Ratnayaka K, Lederman RJ. MRI catheterization in cardiopulmonary disease. Chest. 145, 30-36 (2014).
- Buecker A. Safety of MRI-guided vascular interventions. Minim. Invasive. Ther. Allied. Technol. 15, 65-70 (2006).
- Kos S, Huegli R, Hofmann E et al. Feasibility of real-time magnetic resonanceguided angioplasty and stenting of renal arteries in vitro and in Swine, using a new polyetheretherketone-based magnetic resonance-compatible guidewire. Invest. Radiol. 44, 234-241 (2009).
- Marco AND, Riga CV, Hamady M, Cheshire NJW et al. Robotic and navigational technologies in endovascular surgery. Vasc. Manage. 5, 7, E15–E9 (2010).
- Nordon IM, Hinchliffe RJ, Holt PJ et al. The requirement for smart catheters for advanced endovascular applications. Proc. Inst. Mech. Eng. H. 224, 743-749 (2010).
- NRPB. Guidelines on patient dose to promote the optimisation of protection for diagnostic medical exposures. 10, (1999).
- Recommendations of the International Commission on Radiological Protection. Ann. ICRP. 21, 1-201 (1991).
- Brenner D, Elliston C, Hall E et al. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR. Am. J. Roentgenol. 176: 289- 296 (2001).
- Raval AN, Lederman RJ. Real-time magnetic resonance imaging to guide pediatric endovascular procedures. Pediatr. Cardiol. 26, 251-259 (2005).
- Krueger S, Schmitz S, Weiss S et al. An MR guidewire based on micropultruded fiberreinforced material. Magn. Reson. Med. 60, 1190-1196 (2008).
- Wendt D, Thielmann M, Melzer A et al. The past, present and future of minimally invasive therapy in endovascular interventions: A review and speculative outlook. Minim. Invasive. Ther. Allied. Technol. 22, 242-253 (2013).
- Walsh EG, Brott BC, Johnson VY et al. Assessment of passive cardiovascular implant devices for MRI compatibility. Technol. Health. Care. 16, 233-245 (2008).
- Yeung CJ, Karmarkar P, McVeigh ER. Minimizing RF heating of conducting wires in MRI. Magn. Reson. Med. 58, 1028-1034 (2007).
- Condino S, Calabrò EM, Alberti A et al. Simultaneous tracking of catheters and guidewires: Comparison to standard fluoroscopic guidance for arterial cannulation. Eur. J. Vasc. Endovasc. Surg. 47, 53-60 (2014).
- Carey J, Fahim A, Munro M. Design of braided composite cardiovascular catheters based on required axial, flexural, and torsional rigidities. J. Biomed. Master. Res. B. Appl. Biomater. 70, 73-81 (2004).
- Moftakhar P, Lillaney P, Losey AD et al. New generation laser lithographed dual axis magnetically assisted remote-controlled endovascular catheter for interventional MR imaging: In vitro multiplanar navigation at 1.5 T and 3 T versus X-ray fluoroscopy. Radiology. 277, 842-852 (2015).
- Yao W, Schaeffter T, Seneviratne L et al. Developing a magnetic resonance-compatible catheter for cardiac catheterization. J. Med. Dev. 6, 041002 (2012).
- Nordbeck P, Bauer WR, Fidler F, et al. Feasibility of real-time MRI with a novel carbon catheter for interventional electrophysiology. Circ. Arrhythm. Electrophysiol. 2, 258-267 (2009).
- Bönner F, Janzarik N, Smink J et al. MRguided renal denervation: First experience in pigs. J. Cardiovasc. Magn. Reson. 16(suppl1), 33 (2014).
- Kholmovski EG, Coulombe N, Silvernagel J. Real-time MRI-guided Cardiac Cryo-Ablation: A feasibility study. J. Cardiovasc. Electrophysiol. 27, 602-608 (2016).
- Clogenson HC, van Lith JY, Dankelman J et al. Multi-selective catheter for MR-guided endovascular interventions. Med. Eng. Phys. 37, 623-630 (2015).
- Kaiser M, Detert M, Rube MA et al. Resonant 316 marker design and fabrication techniques for device visualization during interventional magnetic resonance imaging. Biomed. Tech. Berl. 60, 89-103 (2015).
- Rube MA, Holbrook AB, Cox BF et al. Wireless MR tracking of interventional devices using phase-field dithering and projection reconstruction. Magn. Reson. Imaging. 32, 693- 701 (2014).
- Eldirdiri A, Courivaud F, Palomar R et al. Catheter tip tracking for MR-guided interventions using discrete Kalman filter and mean shift localization. Int. J. Comput. Assist. Radiol. Surg. 9, 313-322 (2014).
- Kuehne T, Yilmaz S, Steendijk P et al. X Magnetic resonance imaging analysis of right ventricular pressure-volume loops: In vivo validation and clinical application in patients with pulmonary hypertension. Circulation. 110, 2010-2016 (2014).
- Mekle R, Hofmann E, Scheffler K et al. A polymer-based MR-compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). J. Magn. Reson. Imaging. 23, 145–155 (2006).
- Elgort DR, Hillenbrand CM, Zhang S et al. Image-guided and -monitored renal artery stenting using only MRI. J. Magn. Reson. Imaging. 23, 619-627 (2006).
- Spuentrup E, Ruebben A, Schaeffter T et al. Magnetic resonance--guided coronary artery stent placement in a swine model. Circulation. 105, 874-879 (2002).
- Buecker A, Spuentrup E, Schmitz-Rode T et al. Use of a nonmetallic guide wire for magnetic resonance-guided coronary artery catheterization. Invest. Radiol. 39, 656-660 (2004).
- Clogenson HCM, Dankelman J, Dobbelsteen JJ van den. Steerable guidewire for magnetic resonance guided endovascular interventions. J. Med. Dev. 8, 021002 (2014).
- Pushparajah K, Tzifa A, Razavi R. Cardiac MRI Catheterization: A 10-year single institution experience and review. Interv. Cardiol. 6, 335- 346 (2014).
- Brecher C, Emonts M, Brack A et al. New concepts and materials for the manufacturing of MR-compatible guide wires. Biomed. Tech (Berl). 59, 147-151 (2014).
- Wolska-Krawczyk M, Rube MA, Immel E et al. Heating and safety of a new MR-compatible guidewire prototype versus a standard nitinol guidewire. Radiol. Phys. Technol. (2013).
- Basar B, Campbell-Washburn AE, Rogers T, et al. Stiffness-matched segmented metallic guidewire for interventional cardiovascular MRI. J. Cardiovasc. Magn. Reson. 17, 414 (2015).
- Omary RA, Unal O, Koscielski DS et al. Realtime MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters. J. Vasc. Interv. Radiol. 11, 1079-1085 (2000).
- Seldinger SI. Catheter replacement of the needle in percutaneous arteriography; a new technique. Acta. radiol. 39, 368-376 (1953).
- Odman P. The radiopaque polythene catheter. Acta. radiol. 52, 52-64 (1959).
- Sonmez M, Saikus CE, Bell JA et al. MRI active guidewire with an embedded temperature probe and providing a distinct tip signal to enhance clinical safety. J. Cardiovasc. Magn. Reson. 14, 38 (2012).
- Kocaturk O, Kim AH, Saikus CE et al. Active two-channel 0.035'' guidewire for interventional cardiovascular MRI. J. Magn. Reson. Imaging. 30, 461-465 (2009).
- Clogenson HCM. MRI-compatible endovascular instruments-improved maneuverability during navigation. Delft University of Technology, The Netherlands, (2014).